TDL-5M Desktop Large Capacity Refrigerated Centrifuge Operation Guide

1. Open the door cover, install the rotor, and tighten the nut.
2. Put the centrifuge tube after adding the liquid into the rotor. The centrifuge tubes must be placed in even symmetry; close the door.
3. Plug in the power socket and turn on the power switch (the power switch is on the left side of the instrument);
4. Set the rotor number, speed, temperature and time:
In the stop state, the user can set the rotor number, speed, temperature, time, press the SET button, the centrifuge is in the setting state, the stop light is on, and the running light flashes; in the running state, the user can only set Speed, temperature, time, press SET button, the centrifuge is in the setting state at this time, the running light is on and the stop light flashes (press "SET" key in the stop state to cycle between time, temperature, speed and rotor number ; Running status
Press the "SET" key to select between time, temperature and rotation speed cyclically).
(1), set the rotor number: press the "SET" key, when the decimal point in the lower right corner of the digital tube showing the rotor number is lit, it enters the rotor number setting, and then press the "â–²" or "â–¼" key to select the centrifuge this work place There are four kinds of rotors to choose from with the rotor number. Note: The set rotor number should be consistent with the selected rotor, and it cannot be set wrong.
(2) Set the speed: press the "SET" key, when the decimal point in the lower right corner of the last digital tube showing the speed lights up, enter the speed setting, and then press the "â–²" or "â–¼" key to determine the speed of the centrifuge this time. (The maximum speed is 5000r / min).
(3) Set the temperature: Press the "SET" key, when the decimal point in the lower right corner of the last digital tube showing the temperature lights up, enter the temperature setting, and then press "â–²" or "â–¼" to determine the working temperature of the centrifuge.
(4) Set the time: Press the "SET" key, when the decimal point in the lower right corner of the last nixie tube lights up, enter the time setting, and then press the "â–²" or "â–¼" key to determine the centrifuge's working time ( The longest time is 99 minutes), the time is countdown.
(5) When the above four steps are completed, press the "ENTER" button again to confirm the rotor, speed, temperature, and time set above, and then press the "START" button to start the centrifuge. The centrifuge starts to run, and the rotor number, speed, temperature in the centrifuge chamber and remaining centrifugation time are displayed on the panel.
5. When the time of the centrifuge counts down to "0", the motor is powered off and the brake starts after 5 seconds. The centrifuge will automatically stop. When the speed is equal to 0 r / min, the buzzer will sound 15 times and press "RCF" Press to cancel the beep; press the "STOP" key during operation, the centrifuge stops running and the buzzer sounds.
6. During operation, if you want to see the centrifugal force, press the "RCF" key (RCF light is on), the centrifugal force at the current speed will be displayed, and it will automatically return to the running state after 4 seconds.
7. During the centrifugal operation, if you need to check the originally set parameters, press the SET button, the instrument displays the originally set parameters; after 4 seconds (without keyboard response), the instrument will automatically return to the running state.
8. During the operation of the centrifuge, if you need to modify the operating parameters, press the SET button to the decimal point to the position of the parameter to be modified, then press the up key or down key to select the parameter to be modified, and press the approval key to confirm the modified parameter .
9. When the rotor stops, open the door and take out the centrifuge tube.
10. Turn off the power switch, and the centrifuge is powered off.

Corrugated Paper Box common compoents:

                  1.Clay coated news back(CCNB),210g - 450g

                  2.Corrugated paper : A,C,B,E,F

Corrugated Paper Box traits:

             Light weights,good buffer,saves the resources,the surface cardboard layer and the internal corrugated paper can use the different quality paper, the heat insulation is good,reduces the cost,the corrugated carton packaging inside the product quantity mainly by the carton maximum allowable weight divided by the product unit weight to calculate the determination.

Corrugated paper box usage:

            Mainly apply to products outer packaging and mailer box.

                 

Corrugated Paper Box

Corrugated Paper Box,Black Paper Box,Apparel Paper Box,Corrugated Gift Paper Box

DongGuan SanQun Packing Products Co.,Ltd , http://www.sanqunpacking.com

Posted on